The Mensor CPT6100 is a high accuracy pressure transducer that combines high performance and cost effectiveness for OEM and test system applications.
RS-232 or RS-485 allows the CPT6100 to communicate with any MS-DOS compatible computer over the serial port. A 9 pin D-sub connector is provided to simplify the connections to the serial port of the system or host computer.
Proprietary characterization techniques help the CPT6100 achieve an accuracy of 0.010% FS and a precision of 0.003% FS over the operating temperature range. Every CPT6100 is temperature compensated from 15 to 45° C to insure a high level of performance at varying temperatures. The accuracy statement includes linearity, hysteresis, repeatability and temperature errors over the operating range.
The lowest gauge pressure range is 0.36 psig, while the lowest absolute range is 5 psia. Zero and span can be adjusted via the serial interface. There are no other adjustments required by the end user.
Similar Products: CPT6180, CPT6000
Toll Free: 1-800-984-4200
Outside the US:
001-512-396-4200