Hand-Held Thermometer

Model CTH6500

Model CTH65I0 EEX ib II B T4


Applications

  • Calibrations of thermometers
  • Temperature measurements within the scope of quality assurance
  • Measurements in the service and maintenance sector
  • Long-term monitoring and online documentation

 


Special Features

  • High accuracy of 0.03 K with Pt100
  • One- and two- channel version available
  • Connection of various types of probes possible
  • Optionally with Ex-approval

Hand- Held Thermometer CTH6500

Description

 

The all-purpose model CTH6500 hand-held thermometer, for superior mobile temperature measurement, is notable for its precision, flexibility and ease of handling. In addition to Pt100 resistance thermometers, it can also process signals from typical thermocouples. Thus temperatures from -200 ... +1760 °C can be measured.

Through its high accuracy of 0.03 K in ranges from -100 ... +150 °C, this instrument can also be used as a reference instrument in biotechnology, pharmaceutical and food industries. The CTH6500 is thus also ideal for all service and maintenance tasks.

Low-drift measuring amplifiers ensure small measurement errors, while easy-to-use adjustment features considerably simplify adjustments and calibrations:

  • Calibration by code for fast setting of standard probes via
    identification numbers
  • Physical calibration of probe and display at one, two or
    three different temperatures

In this way it is possible to reduce measuring errors to a minimum and ensure a high display accuracy.

Additional fields of application

The instrument has been primarily designed for temperature measurement, though it can, with the appropriate probes, also be used for:

  • humidity measurement with a combined temperaturehumidity
    probe and for
  • flow measurement from 0.1 ... 40 m/s with a vane sensor.
    The calibration and adjustment possibilities above are also
    applicable to these measurement parameters.

Remarks

  • Range:  -200°C to 850°C (Pt 100) ; -200°C to 1760 °C (TC)
  • Uncertainty: 30 mK to 50 mK from -200 °C to 200 °C above 200 °C 0,1% o. r. (Pt 100)
  • Resolution: 0,01 °C to 200 °C then 0,1 °C
  • Sensor type: Pt 100 and TC's K, J, L, N, T, S and R
© 2012 - Mensor Corporation