Software upgrades

|
|
CPC 6000
Software installer package |
Caution: The CPC 6000 software upgrade installer is used to upgrade a CPC 6000's software to a version that is compatible with Intelliscale sensors. The new Intelliscale sensors will work in CPC 6000's with software version 1.25 and higher. A software upgrade is necessary when installing Intelliscale sensors in a CPC 6000 with software version 1.24 and below.
Software instillation instructions are included in the CPC 6000 software installer package; the instruction file is named "installer.doc".
 |
|
CPG 2500
Software installer package |
Caution: This Software Upgrade should not be installed on CPG 2500s with software version 1.X.X (less than 2.0.0). For CPG 2500s with Software version less than 2.0.0 consult the factory for assistance.
A CPG 2500 software update is necessary if:
Software version is 2.4.0 or earlier and there is a need to calibrate the CPC 2500 locally in units other than psi.
CPG 2500s with software versions of 2.X.X can be updated in the field if you have access to a computer with Ethernet capability. The installer program will only install the software update with a crossover Ethernet cable attached to the instruments Ethernet port and your PC. For assistance contact Mensor Customer Service or call 512-396-4200 X 26.

|