The Mensor Model 9415 Multi-Channel Pressure Test System is used for rapid testing of pressure transducers. It can be a single or dual rack configured with two, four, or six control channels. The model 9414 (quad or dual) controllers provide the control channels. For each pair of control channels, there is a Model 9412 pressure tank. The maximum range of this system is 160 psi.
A Model 9418 Remote Transducer Module with up to four Mensor CPT6180 remote transducers is used for measurements at the test device and can have full-scale ranges of up to 160 psi (~11 bar). All internal transducers have an accuracy of 0.010 % of full scale and a precision of 0.003 % of full scale. Remote transducers can have accuracies of either 0.010 % of full scale or 0.01% IS-50 (0.010 % of reading from mid to full scale and 0.005% of full scale under 50 % of full scale). See General Specifications and Ratings section near the end of the manual for sizes.