Extending our Re-Zero Interval to 1 Year

Accuracy and stability are critical in pressure measurement applications. Both are subject to having their own errors, with a common issue in pressure sensor applications being drift. Drift in a pressure sensor is the change in output that occurs independently of the actual pressure change.

Drift occurs in multiple forms, most commonly being zero drift as well as span drift. Along with this, some might see nonlinearity as a drift error, even though it can be considered more of a precision error:

  • Zero drift: The output shift of the pressure reading at an instrument zero value.
  • Span drift: The change in gain between readings across the full span of the measurement.
  • Non-linear drift: Pressure distortion at specific points.

For Mensor instruments, a re-zero has traditionally been recommended for every thirty days to fight this sensor drift. This practice was implemented to preserve product accuracy and consistency, since sensor drift over time and can affect measurement performance if not periodically corrected. With improvements to our sensing technology, we have removed the 30 day re-zero requirement for the barometric ranges of the CPG2500 Precision Digital Pressure Indicator and CPT9000 Precision Pressure Transducer. The need for monthly re-zeroing maintenance on these specific devices is no longer required, as now this correction needs to be done once per year, along with the already 1-year long re-calibration interval.

Mensor products with improved re-zeroing periods will be noted with an asterisk (*) on their labels moving forward.

Re-Zeroing and Its Challenges

In simple terms as explained before, a re-zeroing process is achieved by taking an instrument in a known pressure condition and adjusting the device, so it reads zero, or a specific pressure value within its specified accuracy. This correction removes potential zero drift and ensures subsequent measurements remain properly aligned.

Performing a 30-day re-zeroing process can present operational challenges, as many Mensor instruments and other pressure reading products in the market are used in demanding environments or in applications requiring long-duration testing. In these conditions, having to re-zero once per month can disrupt the overall workflow process, risk data inconsistency and increase the overall maintenance effort. By syncing the re-zeroing need with the overall calibration interval, you greatly reduce the operational impact due to product maintenance.

For barometers specifically, other challenges like having to use a dead-weight tester or other high accuracy pressure reference to perform a re-zero are added to the fact that this process must occur once a month. The reason a deadweight tester or other high accurate reference must be used for a barometer is because barometers are typically not zero based devices. Mensor barometers along with most barometers operate from a range of 8 to17 psi. The 8 to 17 psi operating range is due to the fact that barometers are designed to measure atmospheric pressure, which naturally fluctuates within this range. For most absolute pressure devices, a vacuum pump to pump the instrument down to a very low pressure then a vacuum sensor is used to create an accurate correction.

Deadweight testers are tedious to use as well, since in most scenarios, a device must be sent to a calibration lab for an accurate deadweight correction. While this is a common annoyance, the reason a deadweight tester is used for barometers is because it is one of the only devices accurate enough to be considered a reference for our barometers. Alternative reference technology is impractical due to strict accuracy requirements. For example, a Mensor CPT9000 with a barometric range requires a 0.008% reading accuracy, and since it is a common practice for a reference to be 4 times as accurate as the device it is correcting, the reference device would have to have an accuracy of 0.002% of reading. This level of accuracy is impractical, which is why a deadweight tester is necessary. For barometers, with this new technology, extending the re-zero interval significantly reduces dead-weight and high accuracy burdens by limiting recalibration to once per year.

Summary

New improved Mensor technology has taken the re-zero interval and increased its period to 1 year, matching that of the Mensor product re-calibration interval. The products with this technology are noted with an asterisk (*) on their label and look to ensure that customers can increase their instrument uptime and simplify their workflows. To summarize the benefits of this new Mensor technology, one can expect less downtime to re-zero their barometer on the CPG2500 and CPT9000. They can also expect more accurate and continuous data collection due to the extended re-zero period, and overall, this new period decreases the monthly requirement for maintenance that otherwise can be tedious.